・Low cost of disposables ・Optimized contrast media management by using large containers ・Efficient workflows ・Minimal residual contrast medium due to automatic switchover from empty to full container(CM-Loop)
Focus on hygiene & safety
・SafeConnect with touch protection ・Easy-Click-Cassette approved for 24-hour use ・Reliable protection against retrograde contamination ・Direct injection from the original media containers ・Patient-side air detector protects against air injection ・Closed system
400mL/patient 1~10mL, in increments of 0.1mL >10~100mL, in increments of 0.5mL >100~400mL, in increments of 1mL
Reservoir capacity
Contrast media: max. 2×200mL(MRI), max. 2×250mL (contrast-enhanced mammography) NaCl 1×1,000mL max. or 1×2,000mL max. with a bag and optional IV pole
Adjustable pressure limit
2~14 bar in increments of 0.1bar(29~203 psi in increments of 1 psi)
Air monitoring
3 detectors for the media supply tubing, 1 detector for the Easy-Click-Cassette and patient tubing
Pressure monitoring
1 pressure sensor
User interfaces
Terminal: 12" color TFT touch screen. Injector: 4" color TFT touch screen
Data import and export
USB interface
Data transfer injector - terminal
WLAN(802.11 b/g/n)
Number of boli(phases) per injection program
40max.
Number of injection programs that can be saved
200max.
Pressure curve
Real-time display of the injection pressure curve
Software functions(standard)
CM-Loop: Alternate switchover with two identical contrast media Display of remaining quantity NaCl backup Manual pause, time-controlled pause Timer Elapsed time
Software options
CM-Select: - Careful selection of a second contrast medium (two different contrast media are available simultaneously) - Possibility to change the type of contrast medium on the media slots by means of NaCI rinsing (no need to change the Easy-Click-Cassette)
Injector accessories
Convenience package: Media rod, waste bin and tray
Disposables
Patient tubing 2.5m or 3.2m Easy-Click-Cassette with holder s or holder I, Adapter rings for small bottles